B-70265EN/02 7.UNITS
- 133 -
(b) Then, check that all discharge tubes are continuously
discharging. If a discharge tube stops discharge, turn VR11
of the power supply corresponding to the discharge tube
clockwise until the discharge tube discharges.
(c) Adjust VR11 of the power supply unit so that only one of
the four discharge tubes is in the state present immediately
before flashing by main discharge occurs (main discharge
disappearing in the other three tubes).
(d) Record the value of RFI------------------------------------------(a)
(e) Turn VR11 counterclockwise to adjust the RFI value to the
value of (RFI - electric shutter adjustment (see Table
7.2.2)).
(f) Turn off the discharge start signal (HV ON), then set the
bias command value in PRM 15223 to 2790 (the current
value minus 60).
(g) Check that trigger electrode discharge lasts for five minutes
in the base discharge state. If discharge is turned off, turn
VR11 of the power supply clockwise slowly, and increase
the value of RFI by 4 mA. Turn the discharge start signal
(HV ON) off then back on, then repeat this operation until
discharge is maintained for five minutes.
(h) Turn off the discharge start signal (HV ON), then reset the
bias command value in PRM 15223 to the original value.
(i) When the base discharge state is entered, record the RFI
value.----------------------------------------------------------------(A)
(2) Base discharge adjustment (C2000-E)
(a) Start the oscillator, and wait until the gas pressure is
stabilized in the base discharge state.
(b) Then, check that all discharge tubes are continuously
discharging. If a discharge tube stops discharge, turn VR11
of the power supply corresponding to the discharge tube
clockwise until the discharge tube discharges.
(c) Adjust VR11 of the power supply unit so that all discharge
tubes are in the state present immediately before flashing by
main discharge occurs.
(d) Record the value of RFI------------------------------------------(a)
(e) Turn VR11 counterclockwise to adjust the RFI value to the
value of (RFI - electric shutter adjustment (see Table
7.2.2)).
(f) Turn off the discharge start signal (HV ON), then set the
bias command value in PRM 15223 to 2790 (the current
value minus 60).
(g) Check that trigger electrode discharge lasts for five minutes
in the base discharge state. If discharge is turned off, turn
VR11 of the power supply clockwise slowly, and increase
the value of RFI by 4 mA. Turn the discharge start signal
(HV ON) off then back on, then repeat this operation until
discharge is maintained for five minutes.
(h) Turn off the discharge start signal (HV ON), then reset the
bias command value in PRM 15223 to the original value.