7.UNITS B-70265EN/02
- 136 -
NOTE
1 RFI value recording and adjustment
If the CNC monitor is located in a place distant from the
oscillator, and you have no assistant, record and adjust the
RFI monitor voltage value on the CN18 connector for RFV
and RFI monitoring on the power supply unit instead of the
RFI value displayed on the CNC screen. After the
adjustment, however, record and check the RFI value by
using the monitor indication on the CNC screen. For the
CN18 connector for RFV and RFI monitoring on the power
supply, use the conversion formula presented in Note 2 of
Table 7.2.2.
2 Alarm processing after modification of Intra-tube pressure
at oscillation time and bias command setting
After modification of parameter No. 15242 and No. 15243
(intra-tube pressure at oscillation time) and parameter No.
15223 (bias command), take the actions below.
(1) If an alarm is issued, indicating that the parameters have
been modified, reset the alarm, and turn on the
discharge start signal (HV ON).
(2) If an alarm requesting that the CNC power be turned off
is displayed, and purge operation is started, turn off the
power to the NC upon completion of purge operation,
and restart.
Table 7.2.2 Settings for power supply adjustments
C1000-E C2000-E C4000-E C5000-E C6000-E
S2 setting B201 B201 B201 B201 B201
S5 setting SHORT USUAL USUAL USUAL USUAL
Electric shutter adjustment 100mA 180mA - - -
Maximum command value PRM.15307 PRM.15306 PRM.15307 PRM.15307 PRM.15307
Maximum power supply output
DCI×DCV=
7600
±50W
7800
±50W
8300
±50W
8700
±50W
8000
±50W
Maximum base output 0W 0W 0W 0W 0W
Maximum RFV <4600V <4600V <4600V <4600V <4600V
Pulse test command Maximum PRM.15307 Maximum Maximum Maximum
Pulse test time >30min >30min >30min >30min >30min
NOTE
1 Power supply output = DCI×DCV,
DCI= (Voltage between DCI-N and DCI-P of CN17)×5,
DCV= (Voltage between DCV-N and DCV-P of CN17)×30
2 RFI= (Voltage between RFI-N and RFI-P of CN18)/2,
RFV= (Voltage between RFV-N and RFV-P of CN18)×1000