10.REPLACEMENT PROCEDURES B-70255EN/01
- 174 -
NOTE
Wrong cable connections to these terminals can
damage the inverter, and can contaminate the
oscillator and damage the discharge tubes by turning
the turbo blower in the reverse direction.
(10) Install the front cover.
(11) Connect the cable to the connector (CN138).
CAUTION
The parameters of an inverter purchased by FANUC
are set before shipment from the factory. So, usually,
the parameters need not be changed. The inverter
can be replaced using the procedure described
above and can be used without modification.
When parameters need to be checked or modified
for a cause, however, see Section 9.3.