6.TROUBLESHOOTING B-70255EN/01
- 86 -
If there exists any anomaly in the results of the above investigations,
replace the anomalous component.
6.4.6 Electromagnetic Contactor of Exhaust Pump Trips Thermally
Possible causes are listed below. Take appropriate action.
Cause of trouble Solution
1 Anomalous setting of
thermal relay
Ascertain if the setting of the contacter is done
at 5.5 A. If not so, correct it.
2 Too high viscosity of oil 1) Check if the room temperature is lower than 5
degree. If so, raise it.
2)Check the deterioration of oil which causes
the viscosity increase. If so, change it.
Recommended exchange period is 1500 H
depending on the surroundings.
3 Reverse rotation
direction of pump
The normal direction is counterclockwise when
seen from the side of motor fan. If not so,
correct it at the terminal.
4 Excessive oil During pump operation, check the oil level
through the oil check window. If the oil level is
between H and L, the oil level is regarded as
normal.
5 Undesignated oil is used. Use the designated oil, NEOVAC SA-H
synthetic oil manufactured by Matsumura Oil
Co. Japan, available through FANUC.
6 Exhaust filter choked. Replace it. Recommended replacement period
is 3000 hours depending on the surroundings.
7 Loosened terminal of
connecting cable
Fix it tightly.
8 Vane damaged Replace the exhaust pump.
6.4.7 Main Breaker Trips
Possible causes are listed below. Take appropriate action.
Cause of trouble Solution
1 Damaged circuit breaker Check for loose terminals. If the circuit breaker
is burned black, replace it.
2 Beyond rated current Check whether the power magnetics cabinet in
the oscillator is abnormal. Turn off the circuit
breakers connected to the power supply and
high-frequency inverter sequentially to
determine the unit that caused excessive
current.