6.TROUBLESHOOTING B-70255EN/01
- 72 -
Alarm output condition
This alarm is issued when the actual gas pressure deviates by ±100 or more
from the set level in LRDY or a later step. Monitoring, however, is not
performed during the time from HV OFF until LRDY.
Related parameters
DGN.905 Actual gas pressure
Cause of trouble Solution
1 Anomaly of parameter
setting of gas pressure
control
Check whether the values of PRM NO.
15000/bit1,15244,15245,15246 are set as
indicated in the attached data sheets. If a
different value is set, set the value specified in
the data sheets.
2 Leakage in circulating
system
1) Check whether any joints in the gas
circulating system are loose. If a loose joint is
found, tighten the joint firmly.
2) Check whether any components of the gas
circulating system are cracked. If a cracked
component is found, replace that component.
3) Check whether there is a hole in a discharge
tube. If a hole is found in a discharge tube,
replace the discharge tube.
3 Stopped turbo blower Check the turbo blower and high-frequency
inverter.
4 Abnormal supply laser
gas pressure setting
Adjust the secondary pressure at the regulator
on the gas cylinder so that the pressure of laser
gas supplied to the laser oscillator is 0.1 to 0.2
MPa (rating) as measured at the entry of the
oscillator. If the distance between the gas
cylinder and the oscillator is larger than 5 m, it is
necessary to set the pressure slightly higher.
5 Anomaly of pressure
sensor
Replace the pressure sensor.
6 The gas flow control
valve is closed.
Adjust the gas flow control valve to supply gas
at the specified flow rate.
7 Anomaly of gas control
unit
If the above checks do not reveal any
abnormality, replace the pressure controller.
ALM No.4079
Press the RESET key
If the emergency stop button is pressed after the discharge start
ready condition (LRDY) is set up, the shutter is closed,
discharge stops, and the LRDY condition is resumed, then this
alarm is displayed. To reset the alarm, remove the cause of
trouble, release the emergency stop button, and press the reset
key on the operator's panel.
Alarm output condition
This alarm is issued when an emergency stop occurs in LRDY or a later step.
Related parameters
DGN.960#1 *ESAL Emergency stop processing signal