B-70255EN/01 APPENDIX E.CONTROL SEQUENCES IN LASER OSCILLATOR
- 253 -
Gas pressure control
200VAC to Laser
power supply unit
PRM.15240
®
PRM.15241
Wait for 120 seconds
PRM.15241+20>
Intra-tube pressure
>PRM.15241-20
WAIT = OFF
LRDY = ON
HV = ON
LRDY = OFF
PSS1 to 8 = Set
PSRDY = ON
RFHV = ON
PTLP = ON
Alarm 2
Monitor treatment start
2
ALM.4081
3
Yes
No
Yes
No
After RUN ON pressure reaches that of PRM15240 and gas pressure
control begins. The gas pressure is changed to the value set in PRM
15240 to the value set in PRM 15241. In addition, when gas pressure
control starts, 200 VAC is supplied to the inverter for the turbo
blower and the power supply unit.
Forty five sec after the beginning of gas pressure control, intra-tube
pressure must be equal to the value PRM15241 within the error of
±20. If not so, alarm 4081 appears.
WAIT = OFF :
Waiting signal for LRDY is off.
LRDY = ON :
Discharge preparation completion signal is on by which HV ON
becomes ready.