3.INSTALLATION B-70315EN/02
- 60 -
3.2.2.6 Antifreezing solution
If the chiller is used in a cold district, it should be provided with an
antifreezing function. When the air temperature falls, the water supply
pump should be kept running. In a very cold district, incorporate a
heater into the chiller to prevent the water temperature to drop too
much. A chiller with a built-in heater is recommended because the
water temperature must reach 20°C or greater to start the oscillator.
If it is necessary to use an antifreezing solution for lack of an
alternative, the following antifreezing solution should be used. Its
concentration should be 30% (usually) or 40% (in an extremely cold
district). Use of an antifreezing solution should be restricted within
four months in winter. The following antifreezing solution is already
added with an anticorrosive; do not use the above-mentioned
anticorrosive.
Product name:
AURORA BRINE
Manufacturer:
Tokyo Fine Chemical Co.
Use:
Refer to the description indicated on the product.